SUMMARY OF CHANGES
Intubation and mechanical ventilation are challenging skills in the austere environment and appropriate consideration must be taken when placing a patient on a mechanical ventilator. Ventilator management is resource intensive and demanding which might not make it appropriate for all tactical situations. If a patient is maintaining their airway and has appropriate oxygenation and ventilation, they should not be intubated by prehospital providers for transport from point of injury (POI) to the first battlefield role of care.
Intubation and providing mechanical ventilation are not without risk. The benefits should outweigh the attributable cost and risk of managing the mechanically ventilated patient, especially within the austere/ battlefield setting. Pathology associated with failure to oxygenate and failure to ventilate will most often require definitive airway interventions and appropriate mechanical ventilator support. At times, it may only serve as a temporizing measure while seeking definitive critical care. Initiation of mechanical ventilation must be guided by clinical suspicion of underlying pathophysiology and clear criteria. A definitive airway is required for effective mechanical ventilation and these patients require sedation – both mechanical ventilation and sedation may make a hypotensive patient more hypotensive and could result in hemodynamic instability. Thoughtful consideration for the requirement of mechanical ventilation must occur. See the JTS Airway Management of Traumatic Injuries and Analgesia and Sedation Management during Prolonged Field Care CPG. In patients with hemorrhagic shock requiring intubation and mechanical ventilation, remember the general principle of “resuscitation before intubation.” Utilize airway adjuncts if the patient’s airway can be maintained adequately during initial resuscitation efforts with blood products. In an under-resuscitated patient, cardiac arrest can occur when induction and paralytic medications are given during rapid sequence intubation.
The intent of this CPG is to provide the non-critical care trained medical personnel with guidance on basic ventilator management. Expanded information on definitions, ventilator terms and modes can be found in Appendix A and B. In the prolonged care setting, telemedicine consultation should be used if possible. Additional JTS CPGs with ventilatory support considerations are Acute Respiratory Failure and Wartime Thoracic Injury.
*Prior to first use, ensure ventilator (as applicable) has been pre-set to utilize parameters of the user’s choice. Some ventilators may be preset to deliver an inspiration time (I-Time) instead of an I:E ratio. Some ventilators default to Volume or Pressure, which would need to be checked prior to use. Failure to appropriately set up and save these ventilator settings may delay use of the ventilator.
1. MODE: Assist Control (AC) or Adaptive Support Ventilation (ASV) -Hamilton-T1 only
Ensure the setting is either Volume or Pressure support. Some ventilators default to different settings.
2. Breaths Per Minute/Respiratory Rate: 14BPM (10-30 range)
3. TIDAL VOLUME (VT): 6ml/kg IBW (4-8ml/kg IBW range)
Quick Reference: (Detailed reference in Appendix C)
Male
66” = ~380cc [min: 255 / max: 510]
69” = ~420cc [min: 283 / max: 566]
72” = ~465cc [min: 310 / max: 621]
75” = ~505cc [min: 338 / max: 676]
Female
60” = ~273cc [min: 182 / max: 364]
63” = ~314cc [min: 210 / max: 419]
66” = ~356cc [min: 237 / max: 474]
69” = ~397cc [min: 265 / max: 530]
4. Fraction of Inspired Oxygenation (FiO2): 21 - 100% (0.21-1.0) (low flow O2 @ 3 LPM (liters per minute) = ~ 40% FiO2 [flowrate on Saros oxygen concentrator])
5. Inspiratory-to-Expiratory (I:E) Ratio: 1:2
6. Positive End-Expiratory Pressure (PEEP): 5 [Range 5-20 cmH20]
7. Pressure Support: 5 [Range 5-20 cmH20] Consider adding if the patient has airway edema or is being ventilated through a narrower diameter ETT.
NOTE: Initial Ventilator settings are based on the patient's ideal body weight (IBW) and clinical condition. However, “basic” or standard starting points are necessary to begin treatment.
1. Set the ventilator to Volume Assist/Control. Consider pressure support if patient meets a clinical need.
2. Set driving mechanics by type of ventilation. Tidal Volume at 4-6 mL/kg IBW (Average adult male is 500 mL).
3. Set rate to maintain an appropriate minute ventilation (VE) of 4-8 L/min (VT x Rate = VE).
4. Set appropriate PEEP. Minimum of 5 cmH2O. Start there and titrate up as needed. In hypoxic patients, consider starting at 10 cmH2O but take extreme caution if also hypotensive as PEEP may worsen BP.
5. Set Fi02. Depending on clinical condition the patient may need high Fi02. Start at 100% and rapidly titrate down to 40% using arterial blood gas and SpO2 data. Do not rely on Sp02 alone (if possible).1
6. Set Inspiratory: Expiratory Ratio (I:E) to 1:2 for most patients. Adjust to clinical conditions.
7. Oxygen requirements can be roughly calculated using Minute Ventilation x FiO2 fraction to determine LPM of pure O2 required (e.g., VE of 6 LPM @ 50% FiO2 (0.5) = 3 LPM bottled or generated O2 required.)
NOTE: Full D cylinder is ~425 liters O2 at 2200psi (i.e. ~141minutes O2 @ 3 LPM requirement above). Calculations should account for discrepancies in tank fill, leaks, dead space, etc. (consider a planning factor of 1.5-2 X calculated requirement).
It is important to test ventilator settings prior to attaching to the patient. A recommendation is to have a test training lung with the equipment. An XL exam glove taped to the circuit is a field expedient test lung. This can ensure the ventilator is appropriately performing.
Determining the need for mechanical ventilation is critical for the effective application of a mechanical ventilation device. Mechanical ventilation has risks, especially in a prehospital environment without diagnostic tools such as CXR or ABG. Clinical acumen is vital to adequately predict a patient who will progress to respiratory compromise. Respiratory compromise requiring ventilator support can be identified early in M-massive hemorrhage, A-airway, R-respiratory, C-circulation, and H-hypothermia (MARCH) assessment.
Inability to adequately oxygenate, ventilate, or guard the airway are indications for mechanical ventilatory support – but the cause must be determined. Indications for intubation must be thoroughly considered as intubation and Positive Pressure Ventilation (PPV) have risks. For example, a patient with increased work of breathing due to hypoxia from pneumothorax will be severely harmed by intubation with the introduction of positive pressure that can expand the pneumothorax and result in lethal tension physiology. Additionally, the bleeding patient who has a very high respiratory rate due to global tissue hypoxemia from hemorrhagic shock may have a full cardiovascular collapse with administration of induction intubation medications and positive pressure ventilation. Thus, it is of utmost importance to ensure that intubation and Mechanical Ventilation (MV) are necessary and critical, with a plan for treating the most likely underlying etiology in place.
NOTE: Apneic patients with adequate circulation and an open airway require immediate assisted ventilation (i.e. bag valve mask [BVM]).
Though ABGs are the standard for managing patients with respiratory compromise, they are not often readily available in the operational environment. Utilization of pulse oximetry (SpO2) and capnography/capnometry (EtCO2) can provide a rapid assessment of a patient’s respiratory status (devices to achieve both are readily available in medical logistics systems).
SpO2 values < 90% are indicative of potential oxygenation issues; however, values can be unreliable due to poor perfusion and altitude.
EtCO2 values >45mmHg are indicative of hypoventilation, especially in the absence of tachypnea. Values <35mmHg indicate hyperventilation. It should be determined why the patient is tachypneic and the underlying cause treated. Mental preparation for mechanical respiratory support should be considered, but hypovolemia/hemorrhage must be treated first. Of note, EtCO2 may not perfectly reflect the arterial CO2 value in patients with significant lung injury and impaired gas exchange.2
A high index of suspicion for sepsis or another cause of metabolic acidosis should be considered in patients with low EtCO2 and hyperventilation.
The inability for a casualty to maintain appropriate SpO2 or EtCO2 values through less invasive measures (airway adjuncts, supplemental O2, etc.) indicates the need for escalating support. For locations with point of care blood analyzers, inadequate blood gas values also indicate the need for respiratory support. Hemorrhagic shock will result in tachypnea and a low EtCO2 . Intubation and mechanical ventilation prior to adequate blood resuscitation will result in circulatory collapse. It is critical to determine why the patient is tachypneic prior to intubation. Ruling out and treating both hemorrhage and tension pneumothorax (PTX) must occur before intubation. Additionally, a simple PTX which could result in moderate tachypnea, positive pressure ventilation can rapidly convert a simple PTX into a tension PTX which is immediately life threatening.
WARNING: Prior to the application of mechanical ventilation and/or initiation of invasive airway devices (e.g., Endotracheal Tube (ETT), extraglottic airways), patients must receive adequate sedation (and paralysis as appropriate/required). Never paralyze a patient who has not received sedation (in the prehospital environment, Ketamine should be first line for sedation) Detailed guidelines for the initiation and maintenance of this sedation can be found in the JTS Analgesia and Sedation Management during Prolonged Field Care CPG.
If respiratory insufficiency amenable to respiratory support is identified during the assessment, the initiation of timely mechanical ventilation can improve casualty outcomes. This can be as simple as BVM support or as complex as a critical care transport ventilator (e.g. Hamilton-T1 or Zoll EMV+), See Appendix D: Impact 754, Appendix E: Zoll EMV+ (731 Series), Appendix F: Hamilton-T1, Appendix G: Save II, Appendix H: Ventway Sparrow. 3,4
All casualties requiring advanced airway management should have a Heat and Moisture Exchanger (HME) attached to the BVM or the circuit tubing. This aids in humidification, heat conservation, reduced infection risk, and comfort/compliance. If used with an EtCO2 cap or device, the HME should be placed above the EtCO2 cap/device to enable more accurate EtCO2 measurements.
CAUTION: BVM respiratory support should be utilized as a temporary measure until mechanical ventilators can be applied. BVMs deliver irregular tidal volume (VT) and respiratory rates.
Although PEEP is helpful in maximizing alveoli recruitment it can also contribute to worsening hypotension in hypovolemic patients and concomitant aggressive resuscitation.
NOTE: Many patients may require increased sedation (and paralysis) for effective ventilation. Patients who are not properly sedated may cause issues with high pressure alarms as they breathe over the ventilator increasing their minute volume and intrathoracic pressures.
Gastric decompression is also a necessary step after securing the definitive airway and initiation of mechanical ventilation.
AIRWAY COMPROMISE OR LOST AIRWAY
If at any time the patient begins to de-saturate or develop respiratory problems, immediately disconnect the ventilator, and manually ventilate the patient with BVM (with PEEP valve if available) and 100% O2 while correcting issues utilizing the following D.O.P.E. algorithm.1
High pressure alarms/Peak airway pressure alarms (Peak pressure >35 cm H2O): Correct problems causing increased airway resistance and decreased lung compliance, including pneumothorax or pulmonary edema. Check the ventilator to ensure the prescribed tidal volume is being delivered. Check for linked/crushed tubing.
Air leaks causing low pressure alarms/volume loss: ETT cuff malfunction/rupture is most commonly identified through loss of VT or audible leak. Assess, correct air leaks in endotracheal tube, tracheostomy cuff, ventilator system; recheck ventilator to make sure prescribed tidal volume is delivered.3
VENTILATOR DYSSYNCHRONY
Ventilator dyssynchrony is a clinical entity in which ventilator gas delivery and patient respiratory mechanics are not matched. Agitation and respiratory distress that develop in a patient on a mechanical ventilator who has previously appeared comfortable represents an important clinical circumstance that requires a thorough assessment and an organized approach. The patient should not always be automatically re-sedated but must instead be evaluated for several potentially life-threatening developments that can present in this fashion.7
Lung hyperinflation (air trapping) and auto-PEEP: Dynamic hyperinflation is associated with positive end-expiratory alveolar pressure, or auto-PEEP. The physiologic effects of air trapping include decreased cardiac preload because of diminished venous return into the chest. This can lead to hypotension and, if severe, to pulseless electrical activity and cardiac arrest. Dynamic hyperinflation can also lead to local alveolar over-distention and rupture. Prevent and manage lung hyperinflation by decreasing tidal volume, changing inspiratory and expiratory phase parameters, switching to another mode, and correcting physiological abnormalities that increase airway resistance.8,9 In an emergency, auto-PEEP from air trapping can be relieved by simply disconnecting the circuit from the endotracheal tube for 3-5 seconds, and then reconnecting.
1. Assess patient’s prior interventions.
2. Assess the following patient’s respiratory status:
3. Assess monitors:
4. Determine cause of ventilation issue and/or ventilator alarm using DOPE mnemonic.
Recently published USAF EMS protocols
Possible causes:
Possible causes:
Source: https://openairway.org/capnography
All patients with an advanced airway requiring mechanical ventilation during transport.
PERFORMANCE/ADHERENCE MEASURES
Number and percentage of patients with an advanced airway that arrive with initial PaCO2 35-45mmHg.
The above constitutes the minimum criteria for PI monitoring of this CPG. System reporting will be performed annually; additional PI monitoring and system reporting may be performed as needed.
The system review and data analysis will be performed by the JTS Chief and the JTS PI Branch.
It is the trauma team leader’s responsibility to ensure familiarity, appropriate compliance, and PI monitoring at the local level with this CPG.
MECHANICS/PHYSIOLOGY OF BREATHING
VENTILATION DEFINITIONS
While in the PH and rotary wing transport environments, it is rare to obtain an ABG – knowing the normal values is important to check a patient prior to Role 2 to Role 3 RW transport. The ABG is extremely useful in trauma resuscitations because the pH, lactate and base deficit will give an overall picture of perfusion.
Volume-targeted Modes
Pressure-targeted Modes
Adaptive Support Ventilation (ASV)
ASV provides intelligent ventilation mode that continuously adjusts respiratory rate, tidal volume, and inspiratory time depending on the patient’s lung mechanics and effort. This is similar to “Auto-Flow” or other like settings on different brands of ventilators. Currently only available on the Hamilton-T1.
Continuous Positive Airway Pressure (CPAP)
CPAP Is a type of positive airway pressure, where the air flow is introduced into the airways to maintain a continuous pressure to constantly stent the airways open, in people who are breathing spontaneously.1 CPAP provides constant pressure above that of the atmospheric pressure. Delivered via the upper respiratory tract.
Pressure Regulated Volume Control (PRVC) Ventilation
PRVC ventilation is designed for invasive mechanical ventilation and combines volume and pressure strategies. PRVC delivers a pressure-controlled and tidal volume (VT)–targeted breath using a decelerating flow waveform pattern that allows unrestricted spontaneous breathing with or without pressure support (PS). 2 (may be Impact 731 only). ***
BL (bilevel)
The BL ventilator provides two pressure settings to assist patients breathing spontaneously: a higher inhalation pressure (IPAP) and a lower exhalation pressure (EPAP).NOTE: This feature is only available on the Zoll. ***
***See Acute Respiratory Failure CPG for more detail on advanced ventilator modes/settings.
VENTILATOR ADJUSTABLE SETTINGS
Tidal volume (VT )
The volume of gas, exchanged during a breath and commonly expressed in milliliters. VT is generally set between 4-8ml/kg IBW, to prevent lung over distension and barotrauma.4-8
Ideal IBW
The weight at which tidal volume is calculated against instead of using actual weight. This enables patients to be ventilated in a lung-protective strategy. A quick reference chart can be found in Appendix C. IBW can be calculated manually as follows:
Minute Ventilation (VE)
The average volume of gas entering, or leaving, the lungs per minute, commonly expressed in liters per minute. Also called minute volume. Minute ventilation is the product of VT and RR (respiratory rate). Normal VE is 5 – 10 L/min. 4-8
I:E Ratio
See I:E definition. I:E might need to be adjusted for physiology that requires extended exhalation time. 4-8 For example: Asthma patient may require an I:E of 1:3, 1:4, or 1:5 to allow for more exhalation time.
Flow Rate
Is the velocity at which gas is delivered to the patient, expressed in liters per minute. When the flow rate is set higher, the speed of gas delivery is faster and inspiratory time is shorter. 4-8
Peak Inspiratory Pressure (PIP)
Represents the total pressure that is required to deliver the VT and depends upon various airway resistance, lung compliance, and chest wall factors. It is expressed in centimeters of water (cm H2O).
Sensitivity or trigger sensitivity
Trigger sensitivity Is the effort, or negative pressure, required by the patient to trigger a machine breath, commonly set so that minimal effort (-1 to -2 cm H2O) triggers a breath.1,3 This is usually seen with assist modes of ventilator operation.
Pressure alarms
Pressure alarms ensure that providers are alerted to pressures that fall outside of appropriate ranges and have potential to harm the patient via barotrauma (over-pressure) or under-ventilation (circuit disconnect or under-pressure). Pressures will be determined by placing the patient on the vent for ~1-2 minutes and determining intrinsic peak inspiratory pressure. (Labeled as PEAK on 754 Ventilator (top right); Labeled as peak on Hamilton T1 ventilator (top left); Labeled as PIP on ZOLL EMV+ (731) (right center). Standard alarm settings should be: Need to measure a peak pressure 5 min after setting initial settings.
The source document is at https://archive.org/details/manual_Impact_Uni-Vent_754_Operation_and_Service_Manual
RULE OF 5S (KEEP IT SIMPLE VENT SETTINGS FOR ADULTS)
There are five numbered dials on the front of the Eagle/Impact Vent. Most settings can be set by remembering multiples of five. Do not hook up casualty to the breathing circuit until after the PEEP is set.
IMPACT 754 VENTILATOR PRE-MISSION CHECKS AND TROUBLESHOOTING
Routine Care
Duty Inspection
1. Power Off Checks:
a. Verify inspection is within the calibration date (6-month maintenance cycle).
b. Check the air inlet is clear of obstructions and the filter is in place (Right side of vent).
c. Verify the Gas (“OXYGEN IN” and “AIR IN”) and Patient (“EXHALATION VALVE” and “TRANSDUCER”) connections are clear and tight (Top of ventilator).
d. Verify the “GAS OUT” clear leaf valve is installed and seated (reseat if loose, replace if missing).
e. Inspect green high pressure oxygen hose for cracks, dry rot, threads, black O-ring (replace if damaged).
f. Connect the ventilator to a high pressure oxygen source, turn on the oxygen tank and ensure no leaks are present.
g. Turn off O2 when complete (conduct in environment conducive to hearing leaks).
2. Power On Checks
a. Turn “MODE” (knob 1) to desired setting (AC, SIMV, CPAP). The ventilator will run a SELF-TEST upon set up (ventilator circuit should be disconnected). At this point, (CAL) is not required. If SELF-TEST results in a Calibration Failure, place (1) to CAL until CAL OK is displayed. If the calibration fails, the ventilator must be taken out of service.
b. Check BATT OK
c. Preset ventilator knobs to:
3. Turn OFF.
4.. Make sure the air inlet and gas out ports of the ventilator are protected and covered.
EMERGENCY PROCEDURES
These procedures should be practiced before being performed on live patients.
NOTE: Any known malfunction of the ventilator should be addressed prior to utilization. The following are not for routine use, but for emergencies when alternate ventilatory measures are not available and long term BVM is not practical.
*ALL CAUTIONS, WARNINGS, AND NOTIFICATIONS THAT CORRESPOND WITH THE 754 SCREEN WILL BE IN ALL CAPS*
NOTE: First place the patient on BVM with supplemental oxygen. Second, check oxygen tank volume. Third, check the oxygen lines and connections.
Alternative Methods to Increase Delivered Oxygen Content
1. Oxygen reservoir kit for low pressure supply (Part # 820-0097-15).
2. Oxygen reservoir fashioned from primary circuit and BVM.
a. Connect short portion of main circuit tube to the BVM and to the air-inlet port.
b. Connect BVM oxygen hose to the BVM and regulator.
c. Set regulator to desired setting (~10LPM, but no lower than total minute volume).
3. Oxygen reservoir fashioned from second ventilator patient circuit.
a. Cut/disconnect exhalation valve off of second ventilator circuit.
b. Feed green transducer hose at least ¾ of the way down vent tubing (the goal is to get as close as possible to the air inlet port) and secure in place with tape (do not cover the end of the circuit).
c. Connect the 90 degree/”L” shape fitting of the green transducer line to the oxygen regulator.
d. Connect the opposite end of the vent tubing to the air inlet port.
e. Set regulator on oxygen source to 10 LPM to deliver up to 99% FIO2.
Missing or Damaged “Gas Out” Leaf Valve
Missing Gas Out leaf valve will trigger an alarm, give a DISCONNECT-CHECK CIRCUIT CONNECTIONS, no PEAK value will display, and little to no volume will be delivered to the patient.
WARNING: Occluding “Gas Out” side ports will enable the ventilator to provide full respirations, however, this will eliminate the anti-asphyxia function these ports provide. (Ventilator failure will result in increased resistance in spontaneous respiration) and strict surveillance must be kept on ventilator to ensure any further failure is caught immediately. Patients must immediately be transitioned to BVM in the event of any failure.
Compressor Failure/Alarms (may show CODE 2).
NOTE: This technique will transition the ventilator to using oxygen pressure instead of the compressor to gather drive ventilation and may hasten oxygen usage.
Battery Failure
1. Turn on the ventilator and ensure that the machine is functional and the battery is charged.
2. Attach the ventilator tubing and O2 tubing to the machine.
3. If the patient is a transfer on a vent, maintain ventilator settings from the medical treatment facility.
4. Default Adult Settings
a. MODE: AC (V)
b. BPM: 12
c. I:E 1:3
d. VT: 450
e. PEEP: 5
f. PIP limit: 25
g. FiO2: 21
5. If the patient is "newly" on the ventilator, initial settings should include:
a. MODE: AC
b. BPM/RRate: 14 BPM (10-30 range)
c. TIDAL VOLUME: 6ml/kg IBW (4-8ml/kg IBW range)
QUICK REFERENCE (Male): (Detailed quick reference in Appendix A)
66” = ~380cc [min: 255 / max: 510]
69” = ~420cc [min: 283 / max: 566]
72” = ~465cc [min: 310 / max: 621]
75” = ~505cc [min: 338 / max: 676]
d. FiO2 : 21 - 100% (0.21-1.0) (low flow O2 @ 3 LPM = ~ 40% FiO2 [flowrate on Saros oxygen concentrator])
e. I:E Ratio: 1:2
f. PEEP: 5 [Range 5-20]
6. Monitor waveform on the machine and visually inspect the patient to ensure no “breath stacking.” If this occurs, a high-pressure alarm may sound. However, if breath stacking is suspected even in the absence of alarm – disconnect tubing and allow exhalation and decrease the I:E if possible from 1:2 to 1:4.
CAUTION: The Hamilton-T1 as fielded by the Department of Defense DOES NOT support noninvasive ventilation (NIV). In environments where there is a risk of patients requiring NIV, a different ventilator should be fielded.
2. Perform pre-operational checks.
a. Click PreOp Checks on main page.
b. Conduct Tightness Test.
c. Conduct Flow Sensor Test.
**Do not attach a patient to the ventilator without conducting both tests.**
3. Select modes
a. Input patient Sex and Height (this calculates all alarm values and “normal” ranges. Do not bypass this step)
b. Touch Modes to change ventilator mode.
c. Select ASV.
4. Select settings
a. Set Tidal Volume (4-8 ml/kg IBW) or Pressure Support (not to exceed 30 mmHg).
NOTE: This ventilator is “PEEP Compensated” which means when in Pressure Support mode if your Pressure Support is 20 and your PEEP is 10, your settings are actually 30 over 10. If you want 20 over 10, you need to see the Pressure Support at 10 and the PEEP at 10.
**This can get confusing, seek direction from medical control if necessary.**
b. Set appropriate rate for age group.
c. Set Fi02 (21 to 100%).
d. Set PEEP (5 to 20).
e. Adjust I:E Ratio as necessary.
5. Set Flow Trigger (0.5 to 5). Press the start ventilation button prior to connecting patient.
6. Once the patient is on the ventilator, you may have to adjust alarm parameters
NOTE: This device is intended for short term or transport use, and is not intended to replace other ventilators available. Consider transitioning the patient to another ventilator (731, 754, or Hamilton T1) as soon as feasible.
NOTE: The Ventway Sparrow is a POI and Transport Ventilator with both Pressure and Volume control modes. The ventilator is applicable to adults and pediatric patients weighting >5 kg (11lb.)
1. Power on. Press the on button for three seconds on the rear panel.
2. Connect tubing: Tubing will only attach one way. Ensure any patient circuit is being attached on the dark grey side of ports, as pictured above. Ensure all connections of the circuit are tight; loose connections may cause low pressure or disconnect alarm. Do not connect patient to tubing until confirmation of vent settings is made.
3. Selecting Screen options. To navigate between the screen options, turn the control knob on the left side of the device. When the desired option has been marked by positioning the marker on its location, press the knob to select the option.
4. Editing Fields. While turning the control knob, fields that can be modified are highlighted. To edit a field, press the control knob when positioned on the field. The field will change color. Rotate the control knob to view different values for the field and press the knob to select a value.
NOTE: When the field changes to red, it means that the selection exceeds the normal setup related to the patient weight or type.
5. Starting Ventilation. Select “Start Vent”. Patient Weight will be in Kg. Initial weight will be for pediatric patients, and you will need to scroll to find adult weight parameters.
NOTE: Setting patient weight will automatically set all ventilation parameters.
NOTE: Selecting “new patient” will clear all previous ventilator settings.
6. Ventilation mode. Ensure patient meets criteria for ventilation management. Select appropriate mode.
7. Ventilation parameters.
a. BPM/RRate: 14BPM (10-30 range)
b. TIDAL VOLUME: The sparrow auto-defaults VT based on ideal body weight you have selected on the prior screen. If those volumes do not work, manually adjust the VT to: 6ml/kg IBW (4-8ml/kg IBW range)
QUICK REFERENCE (Male): (Detailed quick reference in Appendix C)
66” = ~380cc [min: 255 / max: 510]
69” = ~420cc [min: 283 / max: 566]
72” = ~465cc [min: 310 / max: 621]
75” = ~505cc [min: 338 / max: 676]
c. FiO2 : 21 - 100% (0.21-1.0) (low flow O2 @ 3 LPM = ~ 40% FiO2 [flowrate on Saros oxygen concentrator])
d. I:E Ratio: 1:2
e. PEEP: 5 [Range 5-20]
8. Monitor waveform on the machine and visually inspect the patient to ensure no “breath stacking.” If this occurs, a high-pressure alarm may sound. However, if breath stacking is suspected even in the absence of alarm – disconnect tubing and allow exhalation and decrease the I:E if possible, from 1:2 to 1:4.
9. When a high-pressure oxygen source that is connected to an Oxygen Mixer is not available, the Ventway Sparrow ventilator can accept oxygen from a low-pressure oxygen source such as an oxygen enrichment kit connected to a flow meter.
To do this, use an optional low-pressure oxygen enrichment system attached to the ventilator air inlet port through an optional Ventway adapter.
Adjust the “O2 coefficient” parameter on the device through the O2 enrichment screen, which can be found under the VENT. PARAMS screen, so that the coefficient is aligned with the actual FiO2 value given to the patient. Select the "O2 ENRICHMENT" option in the VENT. PARAMS menu. The FiO2 value must be measured with a calibrated external oxygen analyzer. If a high-pressure oxygen source is available, an Inovytec-approved Oxygen Mixer can be used to connect to the air inlet port, delivering between 30% to 95% FiO2 to the patient. Measure FiO2 with a calibrated external oxygen analyzer. The oxygen supply pressure shall be according to manufacturer specifications (usually 40-60 psi).
1. Mechanical Ventilation Equipment
2. Airway Management Equipment
3. Oxygen Supply & Delivery
4. Monitoring & Diagnostics
Vital Sign & Ventilation Monitoring
Arterial Blood Gas (ABG) Management
5. Emergency & Backup Equipment
6. Consumables & Miscellaneous Supplies
This comprehensive list covers mechanical ventilation materials in Role 1-3 settings.
For additional information including National Stock Number (NSN), refer to Logistics Plans & Readiness (sharepoint-mil.us)
DISCLAIMER: This is not an exhaustive list. These are items identified to be important for the care of combat casualties.
The purpose of this Appendix is to ensure an understanding of DoD policy and practice regarding inclusion in CPGs of “off-label” uses of U.S. Food and Drug Administration (FDA)–approved products. This applies to off-label uses with patients who are armed forces members.
Unapproved (i.e. “off-label”) uses of FDA-approved products are extremely common in American medicine and are usually not subject to any special regulations. However, under Federal law, in some circumstances, unapproved uses of approved drugs are subject to FDA regulations governing “investigational new drugs.” These circumstances include such uses as part of clinical trials, and in the military context, command required, unapproved uses. Some command requested unapproved uses may also be subject to special regulations.
Additional Information Regarding Off-Label Uses in CPGs
The inclusion in CPGs of off-label uses is not a clinical trial, nor is it a command request or requirement. Further, it does not imply that the Military Health System requires that use by DoD health care practitioners or considers it to be the “standard of care.” Rather, the inclusion in CPGs of off-label uses is to inform the clinical judgment of the responsible health care practitioner by providing information regarding potential risks and benefits of treatment alternatives. The decision is for the clinical judgment of the responsible health care practitioner within the practitioner-patient relationship.
Consistent with this purpose, CPG discussions of off-label uses specifically state that they are uses not approved by the FDA. Further, such discussions are balanced in the presentation of appropriate clinical study data, including any such data that suggest caution in the use of the product and specifically including any FDA-issued warnings.
With respect to such off-label uses, DoD procedure is to maintain a regular system of quality assurance monitoring of outcomes and known potential adverse events. For this reason, the importance of accurate clinical records is underscored.
Good clinical practice includes the provision of appropriate information to patients. Each CPG discussing an unusual off-label use will address the issue of information to patients. When practicable, consideration will be given to including in an appendix an appropriate information sheet for distribution to patients, whether before or after use of the product. Information to patients should address in plain language: a) that the use is not approved by the FDA; b) the reasons why a DoD health care practitioner would decide to use the product for this purpose; and c) the potential risks associated with such use.